prisma30ST BiLevel-ST with adjustable target volume

prisma30ST is the means of choice for patients who suffer from respiratory insufficiency, e.g., position-dependent hypoventilation and/or COPD, and concurrent SDB (Sleep-Disordered Breathing). Equipped with the unique autoS/T mode with adjustable target volume, autoEPAP and backup frequency, prisma30ST provides comfortable ventilation that quickly leads to lasting treatment success.
Manufacturer : Löwenstein Medical Technology
Country Origin : Germany
To rent or purchase, please dial +91 63096 57730 (or) 93934 50077
Major features:
- Ventilation modes CPAP, APAP, S, T, S/T, autoS/T, aPCV
- Broad pressure range (4 – 30 hPa) and several adjustable parameters for greater flexibility
- Target volume assured in every position
- prisma30ST comes with automatic or permanently set back up frequency and optional adjustable alarms for leaks, low minute volume, low tidal volume and apnea
- SCOPES (indication-specific pre-setting) for COPD and hypoventilation with the option of making adjustments at any time
- Optional autoTrigger (Trigger IN and Trigger EX)
- Adaptive inspiratoy pressure rise (Ramp IN) and expiratory pressure reduction (Ramp EX)
- With SD card and USB port
- Eight clearly distinguished analog signals for PSG feed
- Digital prisma connection (DPC)
- Comfort functions such as softSTART and autoSTART-STOP
- Heated breathing tube prismaHYBERNITE
Technical data:
- Temperature range :
- Operation : +5 °C to +40 °C
- Storage : -25 °C to +70 °C
- Air pressure range : 700 – 1060 hPa
- Electrical output : max. 40 VA
- System interface : 12 V DC max. 10 VA
- Power consumption at : 230V
- Operation (therapy) : 0.11A
- Standby : 0.036A
- Alarms : leaks, disconnection (tube), apnea, low minute volume, low tidal volume
- Mode : CPAP, APAP, S, S/T, T, autoS/T, aPCV
- Mean sound pressure level : about 26.5 dB(A) at 10 hPa
- CPAP/APAP operating pressure range : 4 to 20 hPa
- BiLevel operating pressure range : 4 to 30 hPa
- Trigger Inpiration (Trigger IN) : auto or can be set at 3 levels
- Trigger Expiration (Trigger EX) : auto or can be set at 3 levels
- Pressure rise (Ramp IN) : can be set at 4 levels
- Pressure reduction (Ramp EX) : can be set at 3 levels
Additional Information:
- Dimensions (W x H x D) : 170 x 135 x 180 mm
- Weight : about 1.4 kg